Published in final edited form as: Parkinsonism Relat Disord. 2021 Jun 18;88:102–107. doi: 10.1016/j.parkreldis.2021.06.009

Table 1:

Accuracy and Precision of the ActiGraph GT3X+ device worn on the wrist while walking at comfortable speed over-ground and on a treadmill in Parkinson’s disease and controls.

Absolute Accuracy Absolute Precision Relative Accuracy Relative Accuracy
Mean (SD) Total Step Counts Mean Difference in Step Counts SD of the Mean Difference in Step Counts Mean (SD) Percentage Error n≥5% error n≥10% error n≥25% error Coefficient of Variation
PD (n=29) Over-Ground Walking
Manually Counted 630 (53)
Less Affected Wrist ActiGraph 569 (103) 62 114 9.3 (16.6) 7 6 5 0.08
More Affected Wrist ActiGraph 598 (58) 32 68 4.7 (9.1) 6 5 1 0.04
Treadmill Walking
Manually Counted 663 (63)
Less Affected Wrist ActiGraph 587 (79) 76 86 11.0 (12.0) 15 10 3 0.09
More Affected Wrist ActiGraph 597 (63) 66 66 9.6 (9.1) 18 8 3 0.07
Controls (n=31) Over-Ground Walking
Manually Counted 621 (49)
Dominant Wrist ActiGraph 564 (76) 57 83 8.8 (12.5) 14 9 3 0.07
Non-Dominant Wrist ActiGraph 560 (81) 61 87 9.5 (13.4) 11 10 6 0.08
Treadmill Walking
Manually Counted 647 (58)
Dominant Wrist ActiGraph 546 (101) 1010 102 15.2 (15.4) 20 16 8 0.13
Non-Dominant Wrist ActiGraph 533 (97) 114 113 17.0 (16.5) 21 16 9 0.15