Table 2.

Summarized mass spectrometer analysis conditions for isotope dilution mass spectrometry (IDMS) and plutonium isotopic composition (IC) measurements of the 244Pu Spike reference material

Laboratory LLNL LANL CEA
Instrument (s) Nu-Plasma HR MC-ICP-MS Neptune Plus MC-ICP-MS
IsotopX Phoenix MC-TIMS
Neptune MC-ICP-MS
Detection system (IDMS measurements) 244Pu and 239Pu on Faraday cups 241Pu on axial ion counter
239Pu, 240Pu, 242Pu, 244Pu on Faraday cups
244Pu and 239Pu on Faraday cups
Detection system (IC measurements) 244Pu on Faraday cup
239Pu, 240Pu, 242Pu, 241Pu on ion counters
MC-ICP-MS: 244Pu on Faraday cup
239Pu, 240Pu, 242Pu, 241Pu on ion counters
MC-TIMS: 244Pu on Daly detector
239Pu, 240Pu, 242Pu, 241Pu on ion counters
244Pu on Faraday cup
239Pu, 240Pu, 242Pu, 241Pu on ion counters
Introduction System Cetac Aridus II MC-ICP-MS:ESI Apex Omega
MC-TIMS:Resin bead on Re filament
Scott Double Spray Chamber
Mass Bias CRM CRM 137 IDMS: CRM 128
IC: CRM U005-A
IDMS: IRMM 186
IC: IAEA 8745–01
Quality control CRM CRM 138, CRM 126-A, and IRMM 086 CRM 137 and CRM 138 CRM 137 and CRM 138
Corrections applied Detector Baseline, Instrument Background, Detector Gains, Mass Bias Detector Baseline, Instrument Background, Detector Gains, Mass Bias, Detector dead time Detector Baseline, Instrument Background, Detector Gains, Mass Bias

CRMs for calibration and QC measurements that were not previously cited, are described in [1418]. IAEA 8745–01 is not a CRM but has a Certificate of Isotopic Composition from IRMM specifying the metrologically traceable uranium isotopic ratio values for the material (SMS 7259, 30 May 1997)