. 2021 Mar 10;31(8):5498–5506. doi: 10.1007/s00330-020-07668-x

Table 1.

Parameters used for the non-contrast brain CT acquisitions

Parameter Helical mode (n = 35) Sequential mode (n = 11) Volume mode (n = 2)
Tube voltage (kVp) 120 120 120
Effective tube current time product (mAs) 360 350 140/2251
Rotation time (s) 0.75 1.0 0.5/0.751
Collimation (mm) 0.5 × 40 0.5 × 80 0.5 × 320
Pitch 0.625 n.a. n.a.
CTDIvol (mGy) 52.7 43.2/34.22 16.1/25.51

1In total, two patients were scanned using volume mode. One with a lower mAs and rotation time than the other

2One patient was scanned using the large field-of-view instead of the default medium field-of-view, resulting in a lower CTDIvol