Skip to main content
. 2021 Mar 10;31(8):5498–5506. doi: 10.1007/s00330-020-07668-x

Table 4.

The results found when analysing the scans of only the patients having a foreign body. The scores did not show any significant differences in the observer preferences than those found for the entire case set

Hybrid-IR MBIR DLR DLR vs. Hybrid-IR DLR vs. MBIR
Perceived ratings
Noise 2.87 [2.71–3.04] 2.50 [2.34–2.66] 3.26 [3.09–3.42] p = 0.001 p < 0.001
Sharpness 3.13 [2.95–3.30] 2.53 [2.35–2.70] 3.17 [3.00–3.35] p = 0.733 p < 0.001
Natural appearance 3.23 [3.06–3.40] 2.57 [2.40–2.74] 3.23 [3.06–3.40] p = 1.000 p < 0.001
Grey-white matter differentiation 2.94 [2.76–3.13] 2.57 [2.38–2.76] 3.33 [3.14–3.52] p = 0.005 p < 0.001
Artefacts 2.86 [2.68–3.03] 2.50 [2.32–2.68] 2.81 [2.64–2.99] p = 0.734 p = 0.013
Overall perceived image quality 3.13 [2.95–3.30] 2.53 [2.35–2.70] 3.31 [3.14–3.49] p = 0.141 p < 0.001
Quantitative analysis
Standard deviation in lateral ventricle 6.6 ± 1.1 6.4 ± 0.8 5.7 ± 1.2 p < 0.001 p = 0.02
Signal-difference-to-noise-ratio 1.8 ± 0.4 1.9 ± 0.5 2.4 ± 0.7 p < 0.001 p = 0.01
Standard deviation in fourth ventricle* 6.2 ± 0.3 6.3 ± 1.4 5.1 ± 0.9 p = 0.01 p = 0.02

*Two patients were excluded from these measurements as the fourth ventricle was filled with blood