Skip to main content
. 2021 Jun 29;21(13):4441. doi: 10.3390/s21134441

Table 5.

Regression analysis of the difference (Coefficient (slope and intercept), SE = standard error) together with the 95% confidence intervals (CI), p-values and coefficient of repeatability (CR) of the spatiotemporal parameters between Xsens and the OptoGaitTM.

Parameter Coefficient SE t p 95% CI CR 95% CI
SLo (m), SL1 (m) Intercept 0.282 0.058 4.863 <0.0001 0.168 to 0.397 0.250 0.228 to 0.277
Slope −0.258 0.088 −2.921 0.0039 −0.433 to −0.084
SLo (m), SL2 (m) Intercept −0.629 0.059 −10.577 <0.0001 −0.746 to −0.51 0.119 0.108 to 0.131
Slope 0.805 0.081 9.915 < 0.0001 0.645 to 0.965
SL1 (m), SL2 (m) Intercept −0.831 0.050 −16.598 <0.0001 −0.930 to −0.732 0.318 0.290 to 0.352
Slope 1.004 0.074 13.558 <0.0001 0.858 to 1.150
So (m/s), S1 (m/s) Intercept 2.252 0.080 27.974 <0.0001 2.093 to 2.411 0.232 0.212 to 0.257
Slope −1.618 0.057 −27.957 <0.0001 −1.732 to −1.503
S1 (m/s), S2 (m/s) Intercept −1.726 0.135 −12.702 <0.0001 −1.9943 to −1.458 0.252 0.230 to 0.278
Slope 1.246 0.097 12.734 <0.0001 1.053 to 1.439
So (m/s), S2(m/s) Intercept −0.629 0.059 −10.577 <0.0001 −0.746 to −0.512 0.119 0.108 to 0.131
Slope 0.805 0.081 9.915 <0.0001 0.645 to 0.965