Table 4.
Inter-rater and inter-session reliability analysis for RUSI diaphragm thickness within both manual and specific orthosis device measurement methods during relaxed breathing.
| RUSI Diaphragm Thickness (cm) | Examiner 1 Mean ± SD (95% CI) | Examiner 2 Mean ± SD (95% CI) | ICC(2,1) (95% CI) |
SEM | MDC | p-Value |
|---|---|---|---|---|---|---|
| Manual Probe Fixation | ||||||
| Tins right diaphragm | 0.21 ± 0.07 (0.19–0.24) | 0.20 ± 0.07 (0.18–0.23) | 0.965 (0.931–0.982) | 0.013 | 0.036 | 0.010 † |
| Texp right diaphragm | 0.13 ± 0.04 (0.11–0.15) | 0.11 ± 0.04 (0.10–0.13) | 0.865 (0.675–0.938) | 0.014 | 0.040 | 0.002 * |
| Tins-exp right diaphragm | 0.08 ± 0.05 (0.06–0.10) | 0.09 ± 0.06 (0.06–0.11) | 0.953 (0.906–0.976) | 0.014 | 0.039 | 0.047 * |
| Tins left diaphragm | 0.20 ± 0.07 (0.18–0.23) | 0.20 ± 0.06 (0.17–0.22) | 0.936 (0.876–0.967) | 0.016 | 0.045 | 0.117 * |
| Texp left diaphragm | 0.15 ± 0.05 (0.13–0.17) | 0.13 ± 0.05 (0.12–0.15) | 0.784 (0.571–0.890) | 0.023 | 0.064 | 0.025 * |
| Tins-exp left diaphragm | 0.05 ± 0.04 (0.04–0.06) | 0.06 ± 0.03 (0.04–0.07) | 0.828 (0.669–0.911) | 0.014 | 0.040 | 0.172 * |
| Orthosis Device Probe Fixation | ||||||
| Tins right diaphragm | 0.23 ± 0.07 (0.20–0.25) | 0.22 ± 0.07 (0.19–0.24) | 0.861 (0.731–0.928) | 0.026 | 0.072 | 0.141 * |
| Texp right diaphragm | 0.15 ± 0.05 (0.13–0.17) | 0.14 ± 0.05 (0.12–0.16) | 0.927 (0.847–0.964) | 0.013 | 0.037 | 0.410 † |
| Tins-exp right diaphragm | 0.07 ± 0.05 (0.05–0.09) | 0.07 ± 0.06 (0.05–0.09) | 0.852 (0.713–0.924) | 0.021 | 0.058 | 0.717 † |
| Tins left diaphragm | 0.22 ± 0.07 (0.19–0.24) | 0.20 ± 0.07 (0.18–0.23) | 0.920 (0.839–0.959) | 0.054 | 0.027 | 0.051 * |
| Texp left diaphragm | 0.17 ± 0.05 (0.15–0.19) | 0.16 ± 0.05 (0.14–0.18) | 0.877 (0.762–0.937) | 0.017 | 0.048 | 0.106 * |
| Tins-exp left diaphragm | 0.04 ± 0.05 (0.03–0.06) | 0.04 ± 0.05 (0.02–0.06) | 0.884 (0.776–0.940) | 0.017 | 0.047 | 0.687 * |
Abbreviations: CI, confidence interval; ICC, intraclass correlation coefficient; MDC, minimum detectable change; SD, standard deviation; SEM, standard error of measurement; Tins, maximum inspiration time; Texp, maximum expiration time. p < 0.05 was considered as statistically significant for a 95% CI (in bold). * Student t-test for paired samples was used. † Wilcoxon test for paired samples was used.