Table 4.
Carried out tests applying 200-kPa load on one sensor at a time of the sensing matrix.
| Pressure Value Related to the Sensors’ Insole [kPa] | ||||||||
|---|---|---|---|---|---|---|---|---|
| Pressed sensor | P1 | P2 | P3 | P4 | P5 | P6 | P7 | P8 |
| No sensor | 0.26 | 0.25 | 0.16 | 0.31 | 0.13 | 0.79 | 0.23 | 0.77 |
| Sensor 1 | 207.60 | 9.69 | 2.09 | 1.25 | 2.01 | 2.85 | 1.75 | 0.51 |
| Sensor 2 | 66.62 | 209.60 | 9.02 | 0.95 | 1.71 | 2.01 | 0.59 | 0.47 |
| Sensor 3 | 3.20 | 0.05 | 205.15 | 0.84 | 2.16 | 2.14 | 0.32 | 0.51 |
| Sensor 4 | 0.87 | 0.76 | 3.23 | 206.16 | 1.97 | 1.54 | 0.32 | 0.87 |
| Sensor 5 | 0.11 | 0.81 | 0.71 | 1.91 | 201.28 | 3.22 | 0.95 | 0.40 |
| Sensor 6 | 0.08 | 0.78 | 0.63 | 1.25 | 3.54 | 201.43 | 6.00 | 0.06 |
| Sensor 7 | 0.11 | 0.77 | 0.71 | 0.93 | 0.64 | 4.04 | 202.24 | 0.06 |
| Sensor 8 | 0.16 | 0.83 | 0.89 | 1.17 | 1.66 | 4.18 | 3.80 | 209.71 |