Skip to main content
. Author manuscript; available in PMC: 2021 Jul 12.
Published in final edited form as: J Mech Behav Biomed Mater. 2020 Oct 18;113:104142. doi: 10.1016/j.jmbbm.2020.104142

Table 3.

Maximum values for the numerical risk factors (RFs) for ABS and COMP stents under P1, P2 and P3 loads combinations and the corresponding experimental results.

STENT ABS FEA EXP
Fatigue criteria Output # tests
Load case VM FS BM SWT
P1 1.79 0.91 0.83 0.84 2 Run-out 2
P2 2.79 1.31 1.22 1.25 3 Failure 3
P3 1.94 0.97 0.86 0.90 3 Run-out 3
STENT COMP FEA EXP
Fatigue criteria Output # tests
Load case VM FS BM SWT
P1 1.54 1.00 0.84 0.84 2 Run-out 2
P2 3.40 2.22 2.01 1.31 3 Failure 3
P3 1.87 1.17 1.00 0.85 2 Run-out 2