Table 3b.
Sensitivity and specificity of QuickChaser® Auto SARS-CoV-2 among asymptomatic patients.
Asymptomatic subjects |
Real-time RT-PCR |
||
---|---|---|---|
Positive |
Negative |
||
QuickChaser® Auto SARS-CoV-2 | Positive |
30 |
2 |
Negative | 17 | 491 | |
Sensitivity (%) | 63.8 (48.5–77.3) | ||
Specificity (%) | 99.6 (98.5–100) | ||
Positive predictive value (%) | 93.8 (79.2–99.2) | ||
Negative predictive value (%) | 96.7 (94.7–98.0) |
RT-PCR, reverse transcription polymerase chain reaction.
Data in parentheses are 95% confidence intervals.