. 2021 Jul 14;7:67. doi: 10.1186/s40729-021-00347-6

Table 1.

Studied CBCT devices with corresponding codes and scanning parameters

CBCT devices Code Medium FOV standard Small FOV standard Small FOV high-resolution
FOV (cm) kV mA Voxel size (mm) Exposure time (s) DAP (mGy cm2) FOV (cm) kV mA Voxel size (mm) Exposure time (s) DAP (mGy cm2) FOV (cm) kV mA Voxel size (mm) Exposure time (s) DAP (mGy cm2)
3D Accuitomo 170 A 8 × 8 90 5 0.125 30.8 2310 4 × 4 90 5 0.125 17.5 402 4 × 4 90 5 0.08 30.8 706
Veraview X800 B 8 × 8 100 8 0.125 17.9 2112 4 × 4 100 8 0.125 17.9 596 4 × 4 100 8 0.08 17.9 589
X1 C 8 × 8 90 12 0.15 8.73 773 4 × 4 90 12 0.15 8.73 231 4 × 4 90 12 0.1 11.7 309
NewTom VGi evo D 8 × 8 110 3 0.125 4.32 329 5 × 5 110 3 0.125 4.32 148 5 × 5 110 3 0.1 4.32 148
OP 3D Pro E 8 × 8 90 7 0.2 6 899 5 × 5 90 7 0.125 6 452 5 × 5 90 9 0.08 8.7 775
OP 3D F 6 × 9 95 3 0.2 20 940 5 × 5 95 4 0.125 10 582 5 × 5 95 4 0.08 20 1165
CS 9300 G 8 × 8 90 4 0.18 8 455 5 × 5 90 4 0.2 20 572 5 × 5 90 4 0.09 20 572
ProMax 3D Max H 10 × 10 96 7 0.15 15 668 6 × 6 96 7 0.15 15 657 6 × 6 96 12 0.07 15 1155
Orthophos SL 3D I 8 × 8 85 6 0.16 14 657 5 × 5 85 10 0.16 4.3 214 5 × 5 85 6 0.08 14 421
Spectral Dental CBCT J 9 × 9 110 8 0.2 17 462 6 × 6 110 8 0.2 17 200 6 × 6 110 8 0.07 27 527
Veraviewepocs 3D R100 K 8 × 8 90 8 0.125 9.4 1780 4 × 4 90 8 0.125 9.4 525 * * * * * *
KaVo 3D eXam+ L 8 × 8 120 5 0.125 7.4 500 * * * * * * * * * * * *
Pax-i3D Green M 8 × 8 104 5 0.120 11 409 * * * * * * * * * * * *

FOV Field of view, kV kilovoltage, mA milliamperage, DAP Dose-area product indicated by the manufacturer

*Protocol not available