Table 5.

Comparison of EZ-SARS-CoV-2 RT-PCR assay (EZ) results with TaqPath™ COVID-19 Combo Kit Multiplex Real-Time RT-PCR results for 201 NP samples performed on the same elution in the Cornell COVID-19 Testing Laboratory

ABI 7500 TaqPath positive TaqPath negative QuantStudio 5 TaqPath positive TaqPath negative
EZ positive 92 10 EZ positive 91 9
EZ negative 0 99 EZ negative 2 99
Diagnostic sensitivity 100.00 (100.00-100.00) Diagnostic sensitivity 97.85 (94.90–100.80)
Diagnostic specificity 90.83 (85.41–96.24) Diagnostic specificity 91.67 (86.45–96.88)
Positive predictive value 90.20 (84.43–95.97) Positive predictive value 91.00 (85.39–96.61)
Negative predictive value 100.00 (100.00–100.00) Negative predictive value 98.02 (95.30–100.74)
Kappa 0.94 (0.90–0.97) Kappa 0.93 (0.89–0.97)

Results determined on the ABI 7500 platform are on the left and results determined by the QuantStudio 5 platform are on the right. Estimates are followed by 95% confidence interval in parentheses