Comparison between flap probe-based isothermal amplification and RT-PCR diagnoses (rotavirus, Chlamydia trachomatis, and SARS-CoV-2).




RT-PCR method
Positive Negative Total
The novel flap-probe-based isothermal amplification Target:
Rotavirus (sample 1–50)
Positive 27 0 27
Negative 0 23 23
Total 27 23 50
Target:
Chlamydia trachomatis (sample 51–90)
Positive 21 0 21
Negative 0 19 19
Total 21 19 40
Target:
SARS-CoV-2
ORF1ab gene (sample 91–120)
Positive 12 0 12
Negative 0 18 18
Total 12 18 30
Target:
SARS-CoV-2
N gene (sample 91–120)
Positive 12 0 12
Negative 0 18 18
Total 12 18 30