TABLE 4.
Results of AmpliCoV and specific qPCRs of the INSTAND exercise for quality assurance for SARS-CoV-2 diagnostics.
| Sample | Result qPCR | Ct qPCR | AmpliCoV | Genome coverage, SpotON flow cell (depth ≥ 10) |
| INSTAND-340059 | SARS-CoV-2 | 21.4 | SARS-CoV-2 | 99.6% |
| INSTAND-340060 | OC43 | 25.3 | Negative | |
| INSTAND-340061 | SARS-CoV-2 | 31.3 | SARS-CoV-2 | 95.3% |
| INSTAND-340062 | Negative | Negative | ||
| INSTAND-340063 | SARS-CoV-2 | 24.6 | SARS-CoV-2 | 99.1% |
| INSTAND-340064 | SARS-CoV-2 | 28.2 | SARS-CoV-2 | 94.0% |
| INSTAND-340065 | 229E | 24.7 | Negative | |
| NK | Negative | Negative |