Skip to main content
. 2021 Jun 9;26(3):433–444. doi: 10.5603/RPOR.a2021.0065

Table 4.

Output constancy at different time in a day for 300 MU and 10 × 10 cm2 field size

Energy Time of measurements MR (nC) Output [cGy/MU] Difference (%) Average Standard deviation Coefficient of variation (%)
6MV FFF T1 (10 AM) 34.803 1.0007 Ref 1.0032 0.000876 0.08728
T2 (12 Noon) 34.87 1.0026 0.19
T3 (3 PM) 34.833 1.0015 0.14
T4 (6 PM) 34.81 1.0008 0.01