. 2021 Jul 9;7(7):e07532. doi: 10.1016/j.heliyon.2021.e07532

Table 3.

Inter- and intraday precision and accuracy for the analytes in human plasma.

Analytes Concentration (ng/mL)a Intra-day
Inter-day
Measured concentration (ng/mL) Precision (%) Accuracy (%) Measured concentration (ng/mL) Precision (%) Accuracy (%)
INH 48 45.93 7.18 95.71 45.28 6.91 94.36
120 118.83 3.17 98.98 121.50 4.01 101.28
1920 2003.33 3.62 104.36 1956.67 4.88 101.89
4800 5003.33 3.18 104.22 4874.44 3.24 101.59
RIF 200 207.67 11.68 103.80 189.78 12.11 94.91
500 532.50 6.71 106.49 524.06 7.65 104.83
8000 8831.67 4.38 110.40 8210.56 7.27 102.64
20000 21633.33 7.37 108.22 20200 7.14 100.99
EMB 96 86.68 8.43 90.30 84.85 6.81 88.38
240 236.33 4.17 98.36 239.11 4.69 99.61
3840 4023.33 3.24 104.77 3900 4.19 103.89
9600 10030 4.70 104.58 9650 7.27 100.55
PZA 480 482.67 8.14 100.54 457.78 7.43 95.34
1200 1238.33 2.67 103.22 1226.11 5.07 102.16
19200 19566.67 3.29 101.92 19288.89 2.81 100.46
48000 47816.67 1.80 99.61 47927.78 2.80 99.85
AC-INH 192 201.33 11.63 104.91 188.56 10.21 98.23
480 476.17 5.32 99.20 480.50 5.70 100.13
7680 8001.67 5.29 104.20 7829.44 4.42 101.94
19200 19300 4.16 100.51 19188.89 2.97 99.96
a

The QC samples from top to bottom are the LLOQ, LQC, MQC, and HQC.