Table III.
Maxima from 3 T Siemens MRI field homogeneity measurements for 24 and 35 cm DSV phantoms.
| Unit | DSV (cm) | Prisma |
Vida |
mMR |
|||
|---|---|---|---|---|---|---|---|
| Full shim | Gradshim only | Full shim | Gradshim only | Full shim | Gradshim only | ||
| Field homogeneity (ppm) | |||||||
| pk-pk | 24 | 0.386 | 0.696 | 0.816 | 1.232 | 1.046 | 1.208 |
| 35 | 2.378 | 2.075 | 3.268 | 3.296 | a | a | |
| VRMS | 24 | 0.024 | 0.084 | 0.049 | 0.183 | 0.087 | 0.090 |
| 35 | 0.067 | 0.136 | 0.150 | 0.366 | a | a | |
| SD | 24 | 0.023 | 0.084 | 0.048 | 0.183 | 0.087 | 0.090 |
| 35 | 0.067 | 0.136 | 0.150 | 0.366 | a | a | |
| FWHM | 24 | 0.107 | 0.273 | 0.174 | 0.517 | 0.252 | 0.316 |
| 35 | 0.158 | 0.317 | 0.237 | 1.079 | 0.379 | 0.521 | |
Indicates the measurement was affected by significant signal dephasing.