Published in final edited form as: Med Phys. 2020 Jul 19;47(9):4101–4114. doi: 10.1002/mp.14306

Table III.

Maxima from 3 T Siemens MRI field homogeneity measurements for 24 and 35 cm DSV phantoms.

Unit DSV (cm) Prisma
Vida
mMR
Full shim Gradshim only Full shim Gradshim only Full shim Gradshim only
Field homogeneity (ppm)
 pk-pk 24 0.386 0.696 0.816 1.232 1.046 1.208
35 2.378 2.075 3.268 3.296 a a
 VRMS 24 0.024 0.084 0.049 0.183 0.087 0.090
35 0.067 0.136 0.150 0.366 a a
 SD 24 0.023 0.084 0.048 0.183 0.087 0.090
35 0.067 0.136 0.150 0.366 a a
 FWHM 24 0.107 0.273 0.174 0.517 0.252 0.316
35 0.158 0.317 0.237 1.079 0.379 0.521
a

Indicates the measurement was affected by significant signal dephasing.