Published in final edited form as: Med Phys. 2020 Jul 19;47(9):4101–4114. doi: 10.1002/mp.14306

Table V.

Mean ratios of field homogeneity measurement units from 3 T Siemens MRIs for 24 and 35 cm DSV phantoms.

Ratio DSV (cm) Prisma
Vida
mMR
Full shim Gradshim only Full shim Gradshim only Full shim gradshim only
pk-pk/VRMS 24 16.332 7.809 14.093 6.550 12.473 13.213
35 28.929 14.790 21.464 9.432 a a
pk-pk/FWHM 24 3.488 2.271 3.754 2.216 3.971 3.710
35 14.063 6.990 13.349 3.053 a a
VRMS/FWHM 24 0.214 0.290 0.267 0.338 0.322 0.281
35 0.485 0.432 0.622 0.324 a a
a

Indicates the measurement was affected by significant signal dephasing.