Table 4.

Summary of the recovery test of the proposed sensor and the standard method (ICP-OES) for Cu(II), Co(II), Ni(II), Hg(II), and Mn(II) determination in tap water.

Metal ion Added (mM) Proposed sensor (n=5) ICP-OES (n=3)
Found (mM) %Recovery %RSD Found (mM) %Recovery %RSD
Cu(II) - N.D. N.D.
0.080 0.073 ± 0.021 91 5.69 0.093 ± 0.002 116 0.83
0.400 0.410 ± 0.037 103 5.24 0.427 ± 0.002 107 1.45
0.800 0.756 ± 0.021 95 1.98 0.805 ± 0.009 101 0.60
Co(II) - N.D. N.D.
0.020 0.024 ± 0.004 120 0.83 0.021 ± 0.001 105 4.16
0.040 0.047 ± 0.011 118 2.25 0.040 ± 0.001 100 5.61
0.060 0.067 ± 0.018 112 3.48 0.058 ± 0.001 97 2.82
Ni(II) - N.D. N.D.
0.800 0.856 ± 0.057 107 3.19 0.778 ± 0.002 97 0.10
2.000 2.135 ± 0.082 107 2.68 2.145 ± 0.006 107 0.29
4.000 4.056 ± 0.129 101 2.58 4.033 ± 0.016 101 0.29
Hg(II) - N.D. N.D.
0.008 0.009 ± 0.001 113 1.86 0.008 ± 0.001 100 0.08
0.020 0.022 ± 0.003 110 1.19 0.019 ± 0.001 95 0.30
0.040 0.044 ± 0.003 110 2.25 0.040 ± 0.002 100 0.22
Mn(II) - 0.0047 ± 0.0026 0.0014 ± 0.0002
0.0040 0.0092 ± 0.0012 113 1.61 0.0061 ± 0.0001 117 1.13
0.0060 0.0114 ± 0.0023 112 2.92 0.0078 ± 0.0001 107 1.30
0.0080 0.0139 ± 0.0013 115 1.55 0.0098 ± 0.0002 105 1.04