Skip to main content
. 2021 Jul 17;8:52. doi: 10.1186/s40658-021-00400-8

Table 2.

Proposed specification limits for Gallium-68 multi-centre trials

Sphere diameter (mm) Conventional reconstruction (OSEM5/6.4, OSEM+TOF5/6.4) Advanced reconstruction (PSF5/6.4, PSF800)
SUVmax SUVmean SUVpeak SUVmax SUVmean SUVpeak
37 0.91–1.14 0.68–0.88 0.85–1.04 0.92–1.14 0.71–0.89 0.83–1.06
28 0.87–1.11 0.63–0.81 0.81–1.00 0.90–1.17 0.65–0.88 0.83–1.07
22 0.78–1.11 0.55–0.78 0.70–0.94 0.87–1.19 0.58–0.85 0.76–1.04
17 0.74–0.93 0.50–0.67 0.58–0.69 0.84–1.16 0.51–0.79 0.62–0.84
13 0.56–0.66 0.37–0.47 0.37–0.45 0.65–0.99 0.37–0.63 0.39–0.56
10 0.32–0.49 0.20–0.34 0.21–0.31 0.41–0.59 0.25–0.34 0.24–0.34