Contrast - Soft tissue differentiation |
Diagnostic quality Same as planning CT scan |
Good contrast for large density differences such as bone/ tissue/air. Scatter significantly decreases contrast. |
Good contrast for large density differences such as bone/ tissue/air. Scatter significantly decreases contrast. |
Excellent soft tissue contrast |
No soft tissue differentiation but provides quantitative functional information |
|
**** |
*** |
** |
***** |
*** |
Spatial resolution |
Same as planning CT scan – sub millimeter, can be limited longitudinally |
Same as planning CT scan – sub millimeter, longitudinally typically 1 mm |
Same as planning CT scan – sub millimeter, longitudinally typically 2 mm |
Similar to planning CT scan – sub millimeter, longitudinally typically 1 mm |
Typically few mm in each direction PET, depends on body site |
|
***** |
**** |
**** |
**** |
*** |
Motion artifacts |
Fast scan but motion must be managed to avoid artifacts |
Long scan times prone to motion artifacts |
Long scan times prone to motion artifacts |
Scan times can be long or short - prone to motion artifacts |
Very long scan times, prone to blurring from motion |
Clinical motion management solutions |
Yes |
Yes |
No |
Yes |
No |
|
**** |
** |
** |
*** |
* |
Reconstruction artifacts |
Prone to hardening artifacts from high Z materials or elongated body shape, motion |
Same artifacts as CT, as well as scatter, ring, aliasing, and misalignment artifacts |
Same artifacts as CT, zipper artifacts |
Susceptibility, motion, distortion |
Same artifacts as CT, attenuation correction, motion, CT reconstruction as well as partial volume artifacts |
|
*** |
*** |
*** |
** |
** |
Geometry |
|
|
|
|
|
Anatomy |
Anatomy changes from CT, organ localization |
Anatomy changes, organ localization |
Anatomy changes, organ localization |
Anatomy changes, organ localization |
Metabolic uptake changes |
|
**** |
*** |
*** |
***** |
**** |
FOV limitations |
60 cm FOV |
Up to 50 cm FOV, large FOV results in poor image quality |
Up to 50 cm FOV, large FOV results in poor image quality |
Up to 50 cm FOV |
Up to 70cm FOV |
|
**** |
*** |
*** |
**** |
**** |
Patient position issues |
Same as planning CT scan |
Can affect image quality depending on position on treatment couch |
Can affect image quality depending on position on treatment couch |
Bore size may limit patient position, coil placement may limit use of accessories |
May not be same as simulation setup, PET scan bore size may limit patient position |
|
**** |
*** |
*** |
*** |
**** |
Truncated structures |
Same as planning CT scan |
FOV limitations may truncate structures |
FOV limitations may truncate structures |
FOV limitations may truncate structures and external contour |
Same as planning CT scan |
|
|
|
|
|
|
Tracking organ motion |
Not available during treatment |
Not available during treatment |
Not available during treatment |
Available |
Not available during treatment |
|
* |
* |
* |
**** |
* |
Density |
|
|
|
|
|
HU table management |
Same as planning CT scan |
Can build custom HU table 1–2% accuracy in dose calculation |
MVCT number, similar to HU table, must be monitored at high frequency |
Not available, surrogate needed |
Same as planning CT scan if PET-CT |
|
***** |
***** |
*** |
** |
***** |
Online/Offline ART |
|
|
|
|
|
Modality for Online |
CT on rails |
CBCT |
Tomotherapy able to sum plans and “dose of the day” |
MRI-Cobalt, MRI-Linac |
Under development |
|
*** |
*** |
*** |
**** |
* |
Additional dose to patient |
Up to 3 cGy per scan |
Up to 10 cGy per scan |
Up to 5 cGy per scan |
Not applicable |
Up to 3 cGy whole body plus CT dose |
|
*** |
** |
** |
***** |
** |