Skip to main content
. Author manuscript; available in PMC: 2021 Jul 20.
Published in final edited form as: IEEE Trans Ultrason Ferroelectr Freq Control. 2021 Feb 25;68(3):376–388. doi: 10.1109/TUFFC.2020.3037999

TABLE III.

Percent Errors in Acoustic Output Before and After Inverse Filtering to Compensate for Spatial Averaging

Pressure Parameter ARFI Pulsed Doppler
Before After Before After
pc −34% ± 2% 4% ± 3% −34% ± 2% 0% ± 3%
pr −23% ± 4% −2% ± 7% −21% ± 2% −4% ± 5%
pii −46% ± 5% 5% ± 6% −43% ± 2% 3% ± 2%

Values assume a typical membrane hydrophone with 400 μm geometrical sensitive element diameter. Values are means ± standard errors.