Skip to main content
. Author manuscript; available in PMC: 2021 Jul 20.
Published in final edited form as: Int J Comput Assist Radiol Surg. 2012 Apr 27;7(5):647–665. doi: 10.1007/s11548-012-0682-9

Table 2.

Summary of TRE measurements

Experiment Description TRE (mm) p value
1 Pb glass Without Pb glass 0.83 ± 0.25 0.5
With Pb glass 0.87 ± 0.25
2 Tracker-to-target distance Tracker-on-C setup 0.87 ± 0.25 <0.0001
Conventional setup 1.92 ± 0.71
3 Rotating frame of reference Bench setup 0.70 ± 0.32 0.11
C-arm setup 0.86 ± 0.43