. 2021 Jun 23;22(7):286–296. doi: 10.1002/acm2.13318

Table 2.

CT number in each reconstruction algorithm at different radiation doses.

Radiation dose Contrast Reconstruction algorithm
FBP HIR MBIR DLR
CT number 100% High (340 HU) 333.4 ± 16.3 330.4 ± 8.7 318.5 ± 16.4 337.2 ± 7.7
Moderate (120 HU) 124.1 ± 16.1 121.7 ± 8.1 121.0 ± 6.4 120.1 ± 6.5
Low (−35 HU) −35.4 ± 14.8 −31.6 ± 11 −32.8 ± 8.8 −34.8 ± 10.4
50% High (340 HU) 335.9 ± 33 334.6 ± 14.3 320.3 ± 16.9 342.7 ± 12.4
Moderate (120 HU) 112.8 ± 32.6 118.3 ± 10.7 119.2 ± 7.5 116.6 ± 7.2
Low (−35 HU) −33 ± 17.3 −32.6 ± 7.7 −33.8 ± 6.5 −34.6 ± 6.5
25% High (340 HU) 335.9 ± 45.8 327.7 ± 9.7 319 ± 10.5 334.5 ± 11
Moderate (120 HU) 125.3 ± 36.7 124.2 ± 11.5 126.2 ± 10.7 120.9 ± 8.1
Low (−35 HU) −39.3 ± 34.8 −28.9 ± 8.2 −31.8 ± 6.7 −30.6 ± 7.5

Abbreviations: DLR, deep learning‐based reconstruction; FBP, filtered back projection; HIR, hybrid iterative reconstruction; HU, Hounsfield units; MBIR, model‐based iterative reconstruction.