Table 4.

An example of the stereotactic beam isocentricity assessment for the 6MV FFF beam; mean and one standard deviation of the beam offset (BO) collimated by the Aktina Cone, the ball‐bearing offset (BBO) and the beam CAX deviations from the BB phantom (beam isocentricity ‐ BI) in radial and transverse directions averaged over all gantry and collimator angles.

6MV FFF BO (mm) BBO (mm) BI (mm)
Radial (Mean ± 1SD) −0.09 ± 0.08 0.39 ± 0.56 0.48 ± 0.50
Transverse (Mean ± 1SD) −0.06 ± 0.14 0.01 ± 0.29 0.07 ± 0.30