Table 4.
An example of the stereotactic beam isocentricity assessment for the 6MV FFF beam; mean and one standard deviation of the beam offset (BO) collimated by the Aktina Cone, the ball‐bearing offset (BBO) and the beam CAX deviations from the BB phantom (beam isocentricity ‐ BI) in radial and transverse directions averaged over all gantry and collimator angles.
6MV FFF | BO (mm) | BBO (mm) | BI (mm) |
---|---|---|---|
Radial (Mean ± 1SD) | −0.09 ± 0.08 | 0.39 ± 0.56 | 0.48 ± 0.50 |
Transverse (Mean ± 1SD) | −0.06 ± 0.14 | 0.01 ± 0.29 | 0.07 ± 0.30 |