Table 2:

Intra-scan repeatability and inter-scan reproducibility of coronary flow velocity metrics using proposed triggered GA scheme with shifted binning.

Intra-Scan Repeatability (n=8) Inter-Scan Reproducibility (n=15)
First Scan Repeat ICC P First Scan Repeat ICC P
PSV max (cm/s) 11.8 ± 9.8 13.6 ± 7.7 0.92 <0.01 13.1 ± 8.0 12.6 ± 8.0 0.88 <0.01
TSV max (ms) 79.7 ± 76.9 53.0 ± 10.8 −0.11 0.609 89.4 ± 62.9 89.6 ± 65.8 0.52 0.018
PDV max (cm/s) 22.3 ± 5.7 21.6 ± 5.4 0.98 <0.01 20.3 ± 5.6 20.9 ± 5.8 0.81 <0.01
TDV max (ms) 394.8 ± 35.7 398.6 ± 29.5 0.35 0.164 412.6 ± 45.9 437.7 ± 98.5 0.41 0.054
PSV mean (cm/s) 8.9 ± 7.2 9.5 ± 7.2 0.99 <0.01 9.4 ± 6.1 8.5 ± 6.8 0.84 <0.01
TSV mean (ms) 84.7 ± 66.5 53.0 ± 10.8 −0.12 0.619 83.9 ± 51.6 77.5 ± 43.8 0.66 <0.01
PDV mean (cm/s) 15.3 ± 4.8 15.2 ± 4.7 0.99 <0.01 14.2 ± 4.3 14.4 ± 4.5 0.78 <0.01
TDV mean (ms) 400.0 ± 35.0 393.5 ± 25.0 0.33 0.176 416.7 ± 45.9 440.4 ± 95.2 0.36 0.079
flow (ml/cycle) 0.65 ± 0.29 0.65 ± 0.29 0.98 <0.01 0.64 ± 0.24 0.58 ± 0.24 0.86 <0.01
flow rate (ml/min) 48.0 ± 22.6 48.2 ± 21.5 0.99 <0.01 44.0 ± 18.2 40.6 ± 18.8 0.91 <0.01
area (mm2) 12.2 ± 3.7 12.2 ± 4.1 0.97 <0.01 11.7 ± 3.5 11.1 ± 3.1 0.85 <0.01

Flow velocity metrics are in mean ± standard deviation.

Abbreviations: PSV = peak systolic velocity, TSV = time to PSV, PDV = peak diastolic velocity, TDV=time to PDV, max = maximum velocity in vessel ROI, mean = average velocity in vessel ROI, ICC = intraclass correlation coefficients for absolute agreement