Table 9.

Comparison of the sensitivity and specificity of different rapid, extraction-free molecular test kits/protocols in detecting SARS-CoV.

Name of test kits (manufacturer/country) Detection technology Targets Sensitivity Specificity References
Advanta Dx SARS-CoV-2 RT-PCR assay (Fluidigm) RT-PCR N LOD: 6.25 RNA copies/μL 100% [171]
qScript ® XLT qRT-PCR (Quantabio) RT-PCR N LOD: 6–12 RNA/μL(sensitivity: 85%) Unclear [173]
VereRT™ ZeroPrep™ COVID-19 PCR Kit (Veredus) RT-PCR N LOD: 2 RNA copies/reaction Unclear [172]
XFree™ COVID-19 qRT-PCR test (BioGX) RT-PCR N >90% >90% [170]
FastPlex™ Triplex SARS-CoV-2 Detection Kits RT-PCR N 97.9% 95.7% [174]
SalivaDirect (Yale) RT-PCR N LOD: 1.5 RNA copies/μL 100% [167]
Xpert ® Xpress (Cepheid) RT-PCR N, E LOD: 0.005–0.02 PFU/mL 100% [166]
SARS-CoV-2 SANSURE®BIOTECH Novel Coronavirus (Sansure). RT-PCR N, ORF 1ab 69.9–94.6% 100% [169]
SwabExpress (USA) RT-PCR S, ORF 1b LOD: 2–4 RNA/μL (sensitivity: 100%) 99.4% [168]
PrimeDirect® Probe RT-qPCR Mix (Takara) RT-PCR E, RdRp 55.1–91.9% 88% [169]
PrimeScript®RT-PCR (Takara) RT-PCR E, RdRp 69.6–89.2% 100% [169]
ViroReal® Kit RT-LAMP SARS-CoV-2 (Ingenetix) RT-LAMP ORF 1ab LOD: 100 RNA copies/reaction (95%) 99% [175]
CASSPIT (India) CRISPR/Cas S, ORF 1ab LOD: 100 RNA copies/reaction (~100%) 100% [176]
QSanger™-Covid-19 Assay (Swift Biosciences) Nucleic acid sequencing Unclear LOD: 3200 RNA copies/mL Unclear [177]