Table 3.

NCC between the averaged spectral curves of the different acquisition methods. The curves of the pushbroom, the 41-bands, and the 3×3-VIS camera setup are linear fitted between the single bands. The number of data points indicates the used points for NCC calculation.

Camera comparison Kidney Lung Heart Brain λ-area (nm) Points
Spectrometer versus pushbroom 0.9938 0.9978 0.9957 0.9982 500 to 995 1518
Spectrometer versus 41-bands 0.9707 0.9822 0.9817 0.9820 460 to 970 1540
Spectrometer versus 3×3-VIS 0.8369 0.8397 0.5894 0.8533 400 to 700 850
Pushbroom versus 41-bands 0.9782 0.9841 0.9870 0.9824 500 to 970 1434
Pushbroom versus 3×3-VIS 0.9408 0.9718 0.8975 0.9566 500 to 700 579
41-bands versus 3×3-VIS 0.7378 0.7621 0.7376 0.8250 460 to 700 685