. 2021 Jun 22;14(7):601. doi: 10.3390/ph14070601

Table 2.

Details for process validation of [18F]FE-PE2I produced on a Synthera®+ synthesis module.

Test Specification Batch Results
Batch No. PE2I-200513-2 PE2I-200514-2 PE2I-200520-2
Radioactivity 0.2–20 GBq at EOS 14.1 GBq 14.8 GBq 19.8 GBq
Radiochemical yield (DC) Reported value 13% 14% 20%
[18F]Fluoride ≤5% 0.1% 0.2% 0.2%
Other 18F impurities ≤5% 1.0% 0.9% 2.0%
Radiochemical purity at EOS ≥96% 98.9% 99.0% 97.8%
Radiochemical purity after 6 h ≥90% 94.6% 94.6% 92.5%
FE-PE2I ≤1.00 µg/mL 0.45 µg/mL 0.44 µg/mL 0.28 µg/mL
Specified impurity (RT of 3.4 min) ≤1.00 µg/mL 0.15 µg/mL 0.20 µg/mL 0.13 µg/mL
Total unspecified impurities 1.00 µg/mL n.d. n.d. n.d.
Total FE-PE2I and impurities ≤1.00 µg/mL 0.60 µg/mL 0.64 µg/mL 0.41 µg/mL
Molar radioactivity Reported value 619 GBq/µmol 664 GBq/µmol 1393 GBq/µmol

n.d.—none detected.