Skip to main content
. 2021 Jun 22;14(7):601. doi: 10.3390/ph14070601

Table 6.

Specifications and analytical methods for the quality control of [18F]FE-PE2I.

Test Specifications Analytical Methods
Identification of [18F]FE-PE2I Radioactive half-life: 105–115 min
Gamma spectrum shows only the 511 and 1022 keV peaks. The labeled product corresponds in RT to an authentic reference standard of FE-PE2I
T1/2 measurement using a dose calibrator
Gamma spectrum using a NaI well counter
Product ID using HPLC
Radioactivity 0.2–20 GBq at EOS Dose calibrator
Volume 32 ± 1 mL Visual check
Appearance Clear and colorless solution free from visible particulates or cloudiness Visual inspection
pH 4.0–5.0 pH-meter
Residual Kryptofix < 0.075 mg/mL Color spot test
Free [18F]fluoride ≤ 5% HPLC with a radiodetector
Other 18F-labeled impurities ≤ 5% HPLC with a radiodetector
Radiochemical purity
(at EOS)
≥ 96% HPLC with a radiodetector
Radiochemical purity after 6 h (end of the shelf life) ≥ 90% HPLC with a radiodetector
FE-PE2I content ≤ 1.00 µg/mL HPLC (UV, 220 nm)
Specified impurity (HPLC, RT of 3.4 min) ≤ 1.00 µg/mL HPLC (UV, 220 nm)
Total unspecified organic impurities ≤ 1.00 µg/mL HPLC (UV, 220 nm)
Total FE-PE2I and organic impurities ≤ 1.00 µg/mL(maximum injected dose of 10 µg) HPLC (UV, 220 nm)
Radionuclidic purity ≥ 99.9% (< 0.1% radionuclidic impurities) Gamma spectrum using an HPGe detector after the complete decay of fluorine-18 (minimum 48 h after EOS)
Ethanol 3–7% (w/v) Gas chromatography
Acetonitrile ≤ 273 ppm Gas chromatography
Methanol ≤ 2000 ppm Gas chromatography
DMSO ≤ 3333 ppm Gas chromatography
Microbiology Passes the test for sterility (Ph. Eur.) Test for sterility (Ph. Eur.), filtration method
Bacterial endotoxins ≤ 0.5 EU/mL Quantitative LAL analysis (Ph. Eur.)
Microbiology Bioburden: < 10 CFU/100 mL Membrane filtration and media growth