Table 1.

Stimuli characteristics and signal acquisition parameters for the following auditory electrophysiological testing performed: auditory brainstem response (ABR), middle latency response (MLR), late latency response (LLR), and mismatch negativity (MMN)

ABR MLR LLR MMN
Stimulus Click Click 1,000 Hz “DA” (frequent stimulus)
“DI” (deviant stimulus)
% presentation NA NA NA 80% frequent
20% deviant
Intensity 90 dB HL 70 dB HL 90 dB HL frequent
80 dB HL deviant
Rate 27.7 per second 7.1 per second 1.1 per second
Polarity Alternating
Transducers Insert earphones
Filters 100-3,000 Hz 10-300 Hz 1-30 Hz
Amplification 100 k 50 k 100 k
Montage Ipsilateral array
Electrode placement Inverting (-) right or left mastoid noninverting (+) high forehead ground low forehead Inverting (-) right or left mastoid non-inverting (+) vertex ground low forehead