Skip to main content
. 2021 Jul 13;22(3):985. doi: 10.3892/etm.2021.10417

Table II.

Furnace GTA-100 Varian operating conditions for lead measurements.

Step Temperature (˚C) Time (sec) Flow (liters/min) Gas type Read Signal storage
  40 1.0 3.0 Normal No No
  85 5.0 3.0 Normal No No
  95 40.0 3.0 Normal No No
  120 10.0 3.0 Normal No No
  120 5.0 3.0 Normal No No
  400 5.0 3.0 Alternate No No
  400 1.0. 3.0 Alternate No No
  400 2.0 0.0 Alternate No Yes
  2,100 1.0 0.0 Alternate Yes Yes
  2,100 2.0 0.0 Alternate Yes Yes
  2,100 3.0 3.0 Normal No Yes
  40 2.0 3.0 Normal No No
  40 5.0 0.0 Normal No No