Skip to main content
. 2021 Jul 22;38(7):626–634. doi: 10.1089/ees.2020.0347

Table 1.

Resolution, Regression Equation, Determination Coefficient, Quantification Range, Limit of Quantification, and Limit of Detection of Nine Analytes Using the Method from This Study

Elution order Analyte Resolution (R) Regression equation R2 Quantification range (μg/L) LOD (μg/L) LOQ (μg/L)
1 F 3.05 Y = 0.0071X + 0.1517 0.9986 24.9–1000 7.5 24.9
2 Cl 1.03 Y = 0.0083X + 0.0541 0.9984 14.4–10,000 4.3 14.4
3 NO2 2.92 Y = 0.002X + 0.0082 0.9979 1.5–5000 0.4 1.5
4 NO3 5.51 Y = 0.0122X + 0.0009 0.9981 1.9–10,000 0.6 1.9
5 SO42 2.45 Y = 0.0067X + 0.0903 0.9983 9.5–10,000 2.9 9.5
6 Se (VI) 1.55 Y = 0.0066X + 0.0024 0.9999 0.5–9000 0.2 0.5
7 As (V) 2.51 Y = 0.0023X − 0.009 0.9988 2.1–7000 0.6 2.1
8 ClO4 3.71 Y = 0.0025X − 0.0045 0.9992 0.5–10,000 0.1 0.5
9 Cr (VI) NA Y = 0.0041X + 0.0046 0.9998 0.6–10,000 0.2 0.6

R values >1.5 are baseline resolutions.

Cr, chromium; LOD, limit of detection; LOQ, limit of quantification; NA, not applicable (Cr [VI] was the last analyte in the method run; X, concentration (μg/L); Y, peak area ([μS/cm] × min).