Table 3.
Analyte Accuracy of Quantification Using the Method From This Study
| Elution order | Analyte | Spiked concentration, 2 μg/L (n = 6) |
Spiked concentration, 10 μg/L (n = 6) |
Spiked concentration, 100 μg/L (n = 6) |
|||
|---|---|---|---|---|---|---|---|
| Precision (RSD [%]) | Trueness (recovery [%]) | Precision (RSD [%]) | Trueness (recovery [%]) | Precision (RSD [%]) | Trueness (recovery [%]) | ||
| 1 | F– | NA | NA | 20.3 | 47.5 | 2.2 | 99.5 |
| 2 | Cl– | NA | NA | 12.2 | 90.6 | 0.9 | 96.2 |
| 3 | 0.0 | 95.8 | 2.4 | 97.2 | 1.3 | 97.6 | |
| 4 | 60.7 | 42.6 | 3.1 | 99.1 | 0.9 | 96.3 | |
| 5 | NA | NA | 7.3 | 92.6 | 1.5 | 101.0 | |
| 6 | Se (VI) | 3.2 | 106.4 | 1.3 | 100.8 | 0.8 | 102.2 |
| 7 | As (V) | 0.0 | 102.2 | 0.0 | 101.7 | 0.7 | 107.9 |
| 8 | 0.0 | 104.6 | 1.6 | 105.3 | 0.7 | 103.1 | |
| 9 | Cr (VI) | 5.2 | 98.3 | 1.7 | 100.1 | 0.7 | 98.7 |
NA, not applicable (concentration below limit of detection); RSD, relative standard deviation.