PM2.5 criteria/DQIs defined relative to any instrumentation for regulatory air monitoring requirements in the U.S., EU, and China (Williams et al., 2018; Table B1). Note: A description of the terms and abbreviations are listed in Table S3.

Performance Attributes/DQIs Decision Support

Accuracy/Uncertainty RPDflow: ≤2%
%Diffspecifiedflow: ± 5%, ± 5%
%Diffonepointflow: ± 4% %Diffmultipointflow: ± 2%
Tamb (°C): ± 2, ± 2, ± 2
Pamb (mm Hg): ± 10, ≤ 7.5, ± 7.5
RHamb: ± 5%
Clock/timer (sec): ± 60, ± 20
D50: 2.5 ± 0.2 μm
Collection efficiency: σg = 1.2 ± 0.1
Average flow indication error: ≤ 2% Slope: 1 ± 0.15, 1 ± 0.10
Intercept (μg/m3): 0 ± 10, 0 ± 2
Aerosol transmission efficiency: ≥ 97%
Expanded uncertainty: < 25% in 24-hr averages
Zero level: < 2.0 μg/m3
Zero check: 0 ± 3 μg/m3
Maintenance interval: < 14 days
Bias None
Completeness Completeness (%): 85, ≥90
Detection Limit Detection limit (μg/m3): < 2.0, 2
Tamb resolution: 0.1°C
Pamb resolution: 5mm Hg
Measurement Duration Measurement duration: 60min
Measurement Frequency Flow rate measurement intervals: ≤30s
Measurement Range Concentration range: 0–1000 μg/m3, (0–1000 24-hr avg, 0–10000 1-hr avg μg/m3), 3–200μg/m3
Precision CVconc: ≤5%, ≤ 15%
CVflow: < 2%,≤ 2%, (Avg: ≤2%, Inst.: ≤5%) σ:≤2μg/m3
Precision: < 2.5 μg/m3
RMS: 15%
Response Time None
Selectivity Temperature influence: zero temperature dependence under 2.0 μg/m3, < 5.0% change in min and max temperature conditions
Voltage influence: < 5% change in min and max voltage conditions
Humidity influence: < 2.0 μg/m3 in zero air