Skip to main content
. 2021 Jul 1;17(7):1453–1463. doi: 10.5664/jcsm.9224

Table 2.

Values for different cutoffs of manually edited AHI 4% measured by HSAT and the in-laboratory PM vs PSG.

AHI 4%, Events/h Prevalence Sensitivity Exact 95% CI Specificity Exact 95% CI PPV NPV
LB UB LB UB
HSAT vs PSG ≥ 5 77.9 86.7 74.9 93.7 76.5 49.8 92.2 92.9 61.9
≥ 10 58.4 95.6 83.6 99.2 75.0 56.2 87.9 84.3 92.3
≥ 15 53.2 92.7 79.0 98.1 83.3 66.5 93.0 86.4 90.9
≥ 30 40.3 71.0 51.8 85.1 97.8 87.0 99.9 95.7 83.3
In-laboratory PM vs PSG ≥ 5 76.0 91.2 80.0 96.7 83.3 57.7 95.6 94.5 75.0
≥ 10 58.7 95.4 83.3 99.2 87.1 69.2 95.8 91.3 93.1
≥ 15 53.3 97.5 85.3 99.9 91.4 75.8 97.8 92.9 97.0
≥ 30 40.0 83.3 64.5 93.7 97.8 86.8 99.9 96.2 89.8

Prevalence, sensitivity, specificity, PPV, and NPV for different cutoffs of manually edited AHI 4% from HSAT and the in-laboratory PM vs the PSG. Scoring of hypopneas required ≥ 4% oxygen desaturation according to HSAT and the in-laboratory PM and ≥ 4% oxygen desaturation and/or an arousal according to PSG. AHI = apnea-hypopnea index, CI = confidence interval, HSAT = home sleep apnea test, LB = lower bound, NPV = negative predictive value, PM = portable monitor, PPV = positive predictive value, PSG = polysomnography, UB = upper bound.