Table 4.
Bias, precision, and 95% limits of agreement (adjusted for repeated measures) for 7 days of total sleep time, sleep efficiency, and sleep midpoint as compared between single-channel EEG, actigraphy, and diary.
| α | σ | 95% LoA Lower | 95% LoA Upper | |
|---|---|---|---|---|
| Total sleep time, h | ||||
| EEG vs CSD | −0.67 | 1.03 | −2.72 | 1.39 |
| Actigraphy vs CSD | −0.52 | 0.92 | −2.36 | 1.32 |
| EEG vs actigraphy | −0.13 | 0.84 | −1.81 | 1.54 |
| Sleep efficiency, % | ||||
| EEG vs CSD | −7.07 | 11.68 | −30.43 | 16.29 |
| Actigraphy vs CSD | −6.32 | 10.58 | −27.47 | 14.83 |
| EEG vs actigraphy | −0.55 | 9.97 | −20.49 | 19.39 |
| Sleep midpoint, t | ||||
| EEG vs CSD | −0.10 | 0.35 | −0.81 | 0.61 |
| Actigraphy vs CSD | −0.02 | 0.30 | −0.61 | 0.57 |
| EEG vs actigraphy | −0.08 | 0.27 | −0.62 | 0.47 |
CSD = Consensus Sleep Diary, EEG = single-channel electroencephalography, LoA = limit of agreement, t = time, α = bias, σ = precision.