Skip to main content
. 2021 May 19;59(6):e03161-20. doi: 10.1128/JCM.03161-20

TABLE 6.

Mean output values by sample set and fold change between sample set typesa

Method Measles sera mean output value (min, max) Non-measles sera mean output value (min, max) Fold change
Enzygnost 0.678 (0.079, 1.408) 0.022 (–0.241, 0.185) 31.502
Euroimmun 2.382 (0.329, 4.302) 0.241 (0.002, 1.441) 9.883
Euroimmun Nucleoprotein 2.113 (0.261, 6.290) 0.205 (0.036, 4.270) 10.309
IBL 2.826 (0.761, 4.318) 0.373 (0.074, 1.780) 7.563
LIAISON XL 5.211 (0.270, 13.000) 0.151 (0.100, 1.100) 34.556
Microimmune 2.034 (0.080, 3854) 0.078 (0.021, 1.661) 25.909
NovaLisa 37.498 (4.993, 65.877) 5.179 (0.462, 35.606) 7.241
Serion (activity calculator)b 502.273 (5.395, 2,516.651) 5.005 (0.202, 84.103) 100.354
Serion (OD range)b 1.940 (0.222, 3.477) 0.174 (0.017, 1.347) 11.141
Serion (special case formula)b 1.940 (0.222, 3.477) 0.174 (0.017, 1.347) 11.141
a

Min, minimum; max. maximum.

b

Three methods of sample result determination, using the single set of optical density data from the test plates, were provided in the manufacturer’s IFU. All three methods were evaluated.