Published in final edited form as: J Chromatogr A. 2021 May 28;1651:462294. doi: 10.1016/j.chroma.2021.462294

Table 4.

Inter-day (n = 3 days) recovery, accuracy and precision for validation of the analytical method developed.

Recovery Accuracy Precision
Analyte (Conc.) Spike level day1 day2 day3 inter-day %Deviation %RSD
day1 day2 day3 inter-day day1 day2 day3 inter-day
AA (nM) 100 102 108 104 104 ± 3 3.4 6.4 5.9 5 ± 1 5.9 10 11 9 ± 2
501 93.8 94.5 89 92 ± 2 5.3 8.7 9.4 8 ± 2 4.4 10 4.2 6 ± 3
1002 104 104 105 104 ± 4 3.9 4 5.4 4.4 ± 0.7 3.1 3.8 5 4.0 ± 0.8
PA (nM) 96 96.6 105 84.7 95 ± 8 4.9 6.6 18 10 ± 6 5.1 6.2 16 9 ± 5
561 103 100 101 101 ± 1 3.3 1.5 2.2 2.3 ± 0.7 0.2 1.7 2 1.3 ± 0.8
962 102 106 104 104 ± 2 2.1 6.5 3.8 4 ± 2 0.9 4.2 3.4 3 ± 1
KA (nM) 50 114 117 61 97 ± 26 13.3 11 20 15 ± 4 24 18 12 18 ± 5
100 90.3 107 90.4 96 ± 8 6.4 6.6 6.5 6.5 ± 0.1 5.1 7.4 10 8 ± 2
1003 101 94.3 98.5 98 ± 3 2.8 5.5 2.9 4 ± 1 2.6 4.5 1.6 3 ± 1
QA (nM) 197 110 89.6 116 105 ± 11 4.7 6.1 6.9 5.9 ± 0.9 9.2 14 15 13 ± 3
1077 106 107 121 111 ± 7 8.4 5.8 16 10 ± 4 11 7.3 0.6 6 ± 4
5385 98 103 98 100 ± 2 5 3.7 2.3 4 ± 1 5.9 3.5 3 4 ± 1
TRP (μM) 71 92.4 89.4 87.4 90 ± 2 4.8 6.2 7.3 6 ± 1 6.8 0.9 1.8 3 ± 3
142 103 101 104 103 ± 1 2.5 1 3.1 2.2 ± 0.9 2.1 1.9 1.3 1.8 ± 0.3
282 104 103 104 103.7 ± 0.5 3.8 2.2 3.2 3.1 ± 0.7 3.2 0.8 1.8 2 ± 1
KYN (μM) 4 114 118 123 118 ± 4 11 14 18 14 ± 3 5 0.6 0.8 2 ± 2
8 90 89.6 90.6 90.1 ± 0.4 8.7 9.2 8.1 8.7 ± 0.5 0.9 2.2 3 2.0 ± 0.9
16 98.3 99.4 98.8 98.8 ± 0.5 1.5 3.1 2.3 2.3 ± 0.7 0.5 3.3 2.3 2 ± 1

The assay recovery for the validation was calculated as follows: spiked recovery = 100 × (measured concentration for spiked sample - the measured concentration for nonspiked sample)/spiked value.

The measured concentration of spiked sample = Mean basal concentration (X) + spiked conc.

the measured concentration for nonspiked sample = Mean basal concentration (X).

Mean basal concentration is determined as an unknown concentration value obtained from the calibration curve shown in Fig. S2 and text box A.

Spiked value = added standard concentration.

*

The assay accuracy (% deviation) = 100 × І(measured concentration for spiked sample - the theoretical concentration for spiked sample)І/ the theoretical concentration for the spiked sample. Theoretical concentration for spiked sample = Mean Basal concentration + added standard concentration.