Published in final edited form as: J Surg Oncol. 2021 Apr 18;124(3):271–281. doi: 10.1002/jso.26500

Table 6:

False positives and false negative levels with different thresholds set for Detection Ratios when using PTeye™. A Detection Ratio of 1.2 is the threshold set by the manufacturer of PTeye™ for parathyroid identification.

Detection Ratio
Threshold set in
PTeye™ - 1.2
High
confidence
Medium
confidence
Low
confidence
Total
83 patients (overall) 226 27 83 336
 False Positive at 1.2 5 1 13 19
 False Negative at 1.2 0 0 0 0
19 patients (Initial set) Detection Ratio not recorded 54 2 14 70
 False Positive at 1.2 2 0 1 3
 False Negative at 1.2 0 0 0 0
64 patients (2nd set) Detection Ratio recorded 172 25 69 266
 False Positive at 1.2 3 1 12 16
 False Positive at 1.5 3 0 9 12
 False Positive at 1.8 3 0 6 9
 False Positive at 2.0 2 0 6 8
 
 False Negative at 1.2 0 0 0 0
 False Negative at 1.5 0 0 0 0
 False Negative at 1.8 1 1 0 2
 False Negative at 2.0 1 1 0 2