Skip to main content
. 2021 Jul 28;11:15371. doi: 10.1038/s41598-021-94692-6

Table 2.

Examples of scan parameters with restrictive SAR.

Sequence Case Nsl RPE Rsms TE (ms) TR (ms) SAR (%) Fat. Sup. ON SAR (%) Fat. Sup. OFF
Ep2d_bold (Siemens product) 1.6 mm iso, flip angle = 90 1 64 3 2 20 (partial Fourier = 7/8) 1500 101 34
2 60 3 2 21 1500 97 33
3 76 3 2 21 1500 * 35
cmrr_mbep2d_bold 1.4 mm iso, RF excitation pulse duration = 4 ms, flip angle 90 4 78 3 2 22.6 2030 141 80
5 78 3 3 22.6 2000 101 59

*This case could not be acquired with TR shorter than 1860 ms (because of the extra time required for the fat suppression pulses)’.

The reference amplitude for 1 ms 180° hard pulse: 240 V.