Skip to main content
. 2021 Jul 30;8:57. doi: 10.1186/s40658-021-00405-3

Table 1.

Examined reconstruction protocols (n = 16)

Iterations and subsets SCF
Quantitative reconstructionsa
 Clinical acquisition protocolb 4i/10s 0.41 and 1.10
5i/15s 0.41 and 1.10
24i/10s 0.41 and 1.10
 NEMA acquisition protocolc 4i/10s 0.41 and 1.10
5i/15s 0.41 and 1.10
24i/10s 0.41 and 1.10
Diagnostic reconstructionsd
 Clinical acquisition protocolb 2i/10se 0.41 and 1.10e
 NEMA acquisition protocolc 2i/10s 0.41 and 1.10

aWithout postfiltering,

bSixty projections with 20 s/projection,

cOne hundred twenty projections with 10 s/projection,

dWith postfiltering (Butterworth, cut-off frequency = 0.5 cycles/cm, power = 10)

eInstitutional standard for diagnostic imaging, 2i/10s and SCF = 1.10

SCF scatter weighting factor