Skip to main content
. 2021 Jul 30;8:57. doi: 10.1186/s40658-021-00405-3

Table 3.

Effect of examined parameters on background ACrec and N in comparison to a reference protocola

Parameter of variation ΔACrec pb ΔN pb
(kBq/ml) (%)
SCF: 0.41 + 1.7 < 0.0001 − 6.0 0.02
Iteration set: 4i/10s − 0.1 n.s.d + 22.0 < 0.0001
Iteration set: 5i/15s − 0.1 n.s.d + 35.3 < 0.0001
Iteration set: 24i/10s − 0.2 n.s.d + 68.1 < 0.0001
Acquisition: NEMAc + 0.7 < 0.0001 − 4.1 n.s.d

aClinical standard for diagnostic imaging and reconstruction (60 projections with 20 s/projection, SCF1.10, 2i/10s, Butterworth postfilter)

bANOVA post hoc test

cNEMA acquisition protocol (120 projections and 10 s/projection)

dNot significant, tested by ANOVA

SCF scatter weighting factor, ACrec mean reconstructed activity concentration, N noise