Skip to main content
. Author manuscript; available in PMC: 2021 Jul 31.
Published in final edited form as: Physiol Meas. 2019 Feb 26;40(2):025005. doi: 10.1088/1361-6579/ab033d

Table 5:

Mean and standard deviation, level of agreement, and RSME of the Bland-Altman analysis for the Oxford dataset discriminated by signal quality class.

Quality class Mean (BPM) Standard deviation (BPM) Level of agreement (BPM) RSME (BPM)
Poor 4.72 21.28 [−37.84, 47.28] 21.36
Intermediate 1.08 14.31 [−27.54, 29.69] 14.33
Good 0.52 4.62 [−8.72, 9.76] 4.65
Excellent 0.62 0.12 [0.38, 0.86] 0.63