Table 1.

Determination of sensitivity and precision of the SARS-CoV-2 antigen assay

Inactivated Virus Dilution Calculated input conc. (TCID50/ml)a Calculated input based on Ortho sample (80 µl) Ortho Vitros S/C ratiob (reactive cutoff ≥1.0)c CV%
10−1 4800 384 1150 1180 1160 1.31
10−2 480 38 845 825 826 1.35
10−3 48 4 228 226 229 0.67
10−4 5 0.4 30.2 30.7 31.6 2.30
10−5 0.5 0.04 3.14 3.35 3.3 3.32
10−6 0 0 0.52 0.55 0.51 4.16
10−7 0 0 0.25 0.25 0.26 2.31
10−8 0 0 0.25 0.24 0.23 4.55

a Based on Ortho buffer dilution 4:1.

b Signal/cutoff ratio.

c Each dilution was tested in triplicate.