Table 1.
HRCT technical parameters for the assessment of airways (* kVp: Peak kilovoltage; **mAs: Milliampere seconds; # HU: Hounsfield unit)
Parameter | Value |
---|---|
Volumetric acquisition | - |
Tube potential | 120–80 kVp*; 100 or 80 kVp* to be preferred for small size patients |
Tube current | ≤240 milliampere (mA);≤100 mAs** or effective mAs** |
Slice thickness | ≤1.5 millimetres (mm) |
Gantry rotation time | As short as possible (e.g. 200–500 ms), always ≤1 sec |
Pitch | 1–1.5 |
Reconstruction algorithm | High spatial frequency |
Lung parenchyma windows mean/width | −400 to −700 HU#/>1000 HU # |
Soft tissue windows mean/width | 50/350 HU |