Table 2.

Clinical impact in preprocedural sizing.

Patient no. CT (mm) Freixa Decision SoC Confidence SoC Decision FEops Confidence FEops Implanted
n = 7 (46.7%)     Similar decisions SoC and FEops correlating with implanted device
1 16 × 17 22 20 8 20 9 20
2 17 × 22 25 22 7 22 9 22
8 13 × 19 20 20 8 20 8 20
10 27 × 37 34 34 4 34 7 34
11 19 × 37 34 31 6 31 8 31
13 21 × 36 34 34 6 34 8 34
14 19 × 22 25 25 6 25 8 25
Average       6.4 ± 1.4   8.1 ± 0.7 p=0.003

n = 4 (26.7%)     Different decisions SoC and FEops. FEops correlated with implanted device
3 16 × 18 22 20 8 22 8 22
5 19 × 25 28 25 7 22 7 22
6 14 × 21 22 22 7 20 7 20
7 19 × 26 28 28 7 25 7 25
Average       7.3 ± 0.5   7.3 ± 0.5 p=NS

n = 2 (13.3%)     Similar decisions SoC and FEops, not correlating with implanted device
4 21 × 31 31 31 7 31 8 28
12 22 × 23 28 25 8 25 6 28
Average       7.5   7 p=NS

n = 1 (6.7%)     Different decisions SoC and FEops, none correlating with implanted device
15 14 × 19 22 22 8 20 7 18
n = 1 (6.7%)     Similar decisions SoC and FEops, but no implanted device
9 22 × 31 31 31 7 31 7 NA

CT: computed tomography measurement of landing zone, NS: nonsignificant, SoC: standard of care, and Freixa: sizing based on the sizing table from Freixa et al. [6].