Table 2.
The skin dose-area products (SDAP) and peak skin dose (PSD) as calculated by MC compared to that calculated by the DTS with and without angular correction for a clinical carotid interventional procedure.
| MC | DTS with angular correction | DTS without angular correction | |
|---|---|---|---|
| SDAP (mGy · m2) | 52.8 | 58.1 | 62.6 |
| % SDAP Diff | 0 | 10.0 | 18.6 |
| PSD (Gy) | 2.78 | 2.79 | 2.85 |
| % PSD Diff | 0 | 0.4 | 2.2 |
% Diff = (DTS - MC) / MC × 100