TABLE 3.

Stability results for ManNAc, Neu5Ac, and CMP-Neu5Ac in LQC (surrogate blank matrix) and in HQC (blank matrix)

LQC (75.0 ng/ml) HQC (8067 ng/ml)
ManNAc Deviation (%) Deviation (%)
Room temperature for 18 h 8.7 4.7
Four freeze-thaw cycles 3.7 1.8
Autosampler at 4°C for 67 h 12.4 9.2
Processed sample at 4°C for 71 h 6.5 8.5
Long-term −80°C for 199 days −1.9 1.4
Neu5Ac LQC (75.0 ng/ml) HQC (8975 ng/ml)
Deviation (%) Deviation (%)
Room temperature for 18 h 7.1 −5.7
Seven freeze-thaw cycles −8.7 −13.0
Autosampler at 4°C for 114 h 7.7 3.1
Processed sample at 4°C for 100 h 2.8 −3.1
Long-term −80°C for 198 days 2.1 2.7
CMP-Neu5Ac LQC (30.0 ng/ml) HQC (771 ng/ml)
Deviation (%) Deviation (%)
Room temperature for 21 h 2.0 −10.9
Five freeze/thaw cycles −4.3 −3.5
Autosampler at 4°C for 73 h −2.7 −10.6
Processed sample at 4°C for 73 h 3.7 −9.7
Long-term −80°C for 181 days 0.0 −3.6