Inversion-recovery echo-planar MR in adult brain neoplasia.

S Sheppard 1, P C Davis 1, G Kater 1, J E Peterson 1
PMCID: PMC8338187  PMID: 9504476

Abstract

PURPOSE

A T1-weighted multishot inversion-recovery (IR) echo-planar MR imaging (EPI) sequence was developed to improve intracranial tissue differentiation; its diagnostic utility was compared with that of conventional axial T1-weighted spin-echo and axial T2-weighted turbo spin-echo sequences.

METHODS

Eighteen patients with known or suspected primary or metastatic brain neoplasia were imaged in a 1.5-T unit with IR-EPI sequences. Three observers measured gray/white matter contrast-to-noise ratios and subjectively compared IR-EPI sequences with T1-weighted spin-echo and T2-weighted turbo spin-echo sequences for gray/white matter discrimination, visibility of intracranial and vascular structures, overall lesion conspicuity, size of lesion(s), and presence and severity of artifacts.

RESULTS

Twenty-four lesions (including neoplasia, infarction, treatment-associated encephalomalacia, nonneoplastic white matter signal abnormalities, and basilar artery dolichoectasia) were detected in 12 patients. Basilar artery dolichoectasia was not included in subsequent statistical analysis. Pulsatile flow artifacts were markedly reduced on IR-EPI sequences relative to those on T1-weighted spin-echo sequences. Gray/white matter contrast was greater on IR-EPI images than on T1-weighted spin-echo images. Periaqueductal gray matter, basal ganglia, optic tracts, cranial nerve V, and claustrum were seen better or as well on IR-EPI images as compared with T1-weighted spin-echo images. IR-EPI was more sensitive to magnetic sensitivity effects, with resultant decreased visibility of cranial nerves VII and VIII and the orbital portion of the optic nerves. For noncontrast sequences, lesion conspicuity was better on IR-EPI images than on T1-weighted spin-echo images in 16 (70%) of 23 lesions and was equal on the two sequences in seven (30%) of 23 lesions. Lesion size, including surrounding edema, was greater on IR-EPI images than on T2-weighted turbo spin-echo images in two (9%) of 23 cases and equal in 21 (91%) of 23 cases. Hyperintense foci of methemoglobin were more conspicuous on T1-weighted spin-echo images.

CONCLUSION

Multishot IR-EPI is superior to conventional T1-weighted spin-echo imaging for parenchymal tissue contrast and lesion conspicuity, and is equal to T2-weighted turbo spin-echo imaging in sensitivity to pathologic entities.

Full Text

The Full Text of this article is available as a PDF (817.6 KB).