Skip to main content
. 2021 Jan 27;34(4):631–642. doi: 10.1007/s10334-021-00907-2

Fig. 1.

Fig. 1

Magnet setup. a The Halbach magnet, producing an 0.05 T magnetic field at the center. b The B0 field was measured on a 225 × 225 × 300 mm3 with a gaussmeter connected to a robot. After subtraction of the center frequency and conversion to Hz it was fitted to a basis of second-order spherical harmonics. The field in the center transversal slice of the magnet shows inhomogeneities up to approximately ± 1200 Hz. The dashed lines indicate the position of the inner diameter of the magnet. c Schematic overview of the reconstruction algorithm for iterative CPR and iterative MB. In each iteration, two images are reconstructed from the unshifted and shifted acquisitions. The phase of the resulting images is used to map the ΔB0 field. The ΔB0 map is used as input in the CPR and MB reconstruction in the next iteration